Validation of moving target doses using ArcCHECK Planned Dose Perturbation Algorithm

Authors

  • P K Sharma International Oncology Center, Fortis Hospital, Noida (U.P.), India
  • Roopam Srivastava International Oncology Center, Fortis Hospital, Noida (U.P.), India
  • Laishram Suchitra International Oncology Center, Fortis Hospital, Noida (U.P.), India

DOI:

https://doi.org/10.24297/jap.v9i2.1392

Keywords:

4DCT, ArcCHECK, Respiratory Motion Sim, 3DVH

Abstract

 The aim of the study is to use Planned Dose Perturbation (PDPâ„¢) measurement-guided reconstruction method to estimate dynamic 4D dose and evaluate both 3D dose and DVH changes caused by target motion resulting from respiration. Five patients of Ca lung were selected for the study. Target and Organ at risks were (OARs) delineated on 4DCT data set of each patient. Dose of moving target vs. stationary target were simulated and compared for OARs and target by analyzing 3D dose and DVH(Dose Volume Histogram).There was almost 3.5% higher target maximum doses measured with moving target after applying the target motion trajectory data as compared to the stationary target whereas OARs doses were comparable. The results, clinically signifies the importance of motion management in lung tumors.

Downloads

Download data is not yet available.

Author Biographies

P K Sharma, International Oncology Center, Fortis Hospital, Noida (U.P.), India

Department of Medical Physics

Roopam Srivastava, International Oncology Center, Fortis Hospital, Noida (U.P.), India

Department of Medical Physics

Laishram Suchitra, International Oncology Center, Fortis Hospital, Noida (U.P.), India

Department of Medical Physics

Downloads

Published

2015-06-04

How to Cite

Sharma, P. K., Srivastava, R., & Suchitra, L. (2015). Validation of moving target doses using ArcCHECK Planned Dose Perturbation Algorithm. JOURNAL OF ADVANCES IN PHYSICS, 9(2), 2348–2354. https://doi.org/10.24297/jap.v9i2.1392

Issue

Section

Articles

Most read articles by the same author(s)