Corrosion Behaviour of Metals in Artificial Tear Solution
DOI:
https://doi.org/10.24297/jac.v8i3.4051Keywords:
Corrosion of metals, artificial tear, Ni-Cr, SS 316 LAbstract
Human tear comes in contact with a number of instruments during operation in the eyes. This results in a variety of undesirable effects such as corrosion and malfunction. Corrosion behaviour of five metals, namely, mild steel (MS), mild steel coated with zinc (MS-Zn), Ni-Cr, Ni-Ti super elastic (Ni-Ti.SE), and SS 316 L in artificial tear solution has been studied by polarization study and AC impedance spectra. The study reveals that the decreasing order of corrosion resistance in artificial tear solution is : Ni-Ti SE > Ni-Cr > SS 316 L > MS-Zn > MS. The first three metals are better candidates and the first one is the best candidate for making instruments used in operation in the eyes, in presence of tears.